1. Prior to removing from hose, run handpiece for 10 seconds.
2.
Clean externally with a degreaser/cleaner. NEVER USE SOAP OR WATER! High heat cause soap to crystallize, causing bearings to seize, water causes rust deposits internally!
3.
Apply 3-5 drops of handpiece lubricant to intake side of handpiece (shown above). Please use a low viscosity lube.
IF LUBE DOES NOT EXPEL FROM HEAD CLEAR AND CLEAN REPEAT STEP 3
4. Remove bur, bag handpiece. Make sure to elevate the head, so that sediments may run down the tube rather than rest in bearings where they may cause rust deposits.
Do Not Use Dry Heat, Which May Damage Internal Parts
5. Upon completion of sterilization cycle, open bag to air dry.
Never Leave Handpieces In Sterilizer Over Night |